Quick start guide CellaTemp PA
images is sharp.
Pyrometer with fiber optic cable and separate optic
For pyrometers with fiber optics, the measuring head must be fo-
cused. To activate the laser, press the MODE button on the rear
panel for 2 second. Observe the laser safety instruction. For focal
adjustment, loosen the screw of the measuring head and move
the inner tube in order to obtain a sharp spot on the measuring
area.
6. The pyrometer uses the intensity of infrared radiation for non-con-
tact temperature measurements. It is necessary to configure the
pyrometer to the respective emissivity coefficient of the measuring
object to obtain the exact measuring results. The emissivity coeffi-
cient is to set as following:
► Press [
▲or▼]
► Release [
> The current temperature value is displayed and the new emis-
sivity coefficient is stored.
Make sure that the function of the pyrometer is correct. All param-
eters can be set directly at the pyrometer. Record the setted pa-
rameters.
7. It is advisable to take the checking of the pyrometer in the mainte-
nance plan.
The pyrometers of the PA 1x series are only measuring exact re-
sults in state of terminal stabilization. The break period is approxi-
mately 15 min after starting the supply voltage.
Shipping, packing and disposal
Inspection after shipping
Unpack and inspect the entire shipment immediately upon receipt to
make sure it is complete and undamaged. If the container/package
shows visible signs of damage, please refuse the shipment. If this is not
possible, accept the shipment on the condition that the freight carriers
delivery record is noted with the extent of the damage in order to file a
claim. Should you discover a concealed loss or damage, report it to the
shipper or freight carrier immediately. If the period for filing claims has
expired, you will no longer be able to make any claims for compensation
of damage or loss.
until the desired emissivity is shown.
▲or▼]
.
EN