Varian Turbo-V 2000 ICE Notice De Mode D'emploi page 91

Masquer les pouces Voir aussi pour Turbo-V 2000 ICE:
Table des Matières

Publicité

XX KRPM = indicates the actual theoretical rotational
speed of the pump as a function of the controller output
frequency (range 3 to 33 KRPM).
HJ =
displayed only when the heater jacket is
active.
HS/HL = indicates that the pump is
working in the High speed or
Low speed mode.
After START command, frequency
maximum level, then the
value proportional to the pump rotational speed (about 3
KRPM if the pump is completely stopped).
The pump will accelerate to its normal rotational speed.
During acceleration of the pump or during any
operating condition, it is always possible to select the
other parameters to be displayed pressing the PUMP
CURRENT or the CYCLE NUMBER pushbuttons.
After the run up time and when the normal rotational
speed is reached, the display will be as follows, even
if any previous display selection was made, and the
normal condition has been reached.
N O R M A L
1 2
X X
K R P M
where: XX = indicates the rotational speed (33 KRPM for
high speed, or 26 KRPM for low speed).
output will be at the
frequency will decrease to a
O P E R A T
I O N
H J
H S
Monitor Relay Programming
The display shows:
H .
S
.
S E T T
where: XXKRPM = is the switch point of relay S1 at the
preset turbopump speed, adjustable from 00 to 38 KRPM.
The speed threshold will condition the S1 operation
(see the following cycle diagram] and it is factory-set to
30 KRPM.
The speed threshold will condition the S2 operation
(see the following cycle diagram] and it is factory-set to
23 KRPM.
Pump Shutdown
Press the front panel STOP push-button or remove the
remote signal; the power from the turbopump will be
removed and the pump will begin to slow down.
Power Failure
In the event of a power failure (momentary or long term),
the Turbo-V ICE controller will stop the turbopump and all
the interconnected pumps/devices. The Turbo-V ICE vent
valve device, if used, will vent the turbopump only if the
power failure is longer than the preset delay time. When
power is restored, the Turbo-V ICE controller automatically
restarts the interconnected devices and the turbopump in
the proper sequence
Cycle diagram
87
TECHNICAL INFORMATION
T H R E S H O L D
I
N G ;
X X K R P M
87-900-929-01 (H)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-9448969-9449Turbo-v 2000

Table des Matières