Knick Stratos Pro A2 PH Série Guide Rapide page 18

Masquer les pouces Voir aussi pour Stratos Pro A2 PH Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37
Error Messages
Info text
(is displayed in case of
Error
fault when the Info key is
pressed)
ERR 04
SENSOR FAILURE
ERR 05
CAL DATA
ERR 10
ORP RANGE
ERR 11
PH RANGE
ERR 12
MV RANGE
ERR 13
TEMPERATURE RANGE
ERR 15
SENSOCHECK GLASS-EL
ERR 16
SENSOCHECK REF-EL
ERR 60
OUTPUT LOAD
ERR 61
OUTPUT 1 TOO LOW
ERR 62
OUTPUT 1 TOO HIGH
ERR 63
OUTPUT 2 TOO LOW
ERR 64
OUTPUT 2 TOO HIGH
ERR 69
TEMP. OUTSIDE TABLE
*) Digital sensors (ISM, InduCon, Memosens)
18
Problem
Possible causes
Failure in sensor *
Error in cal data *
ORP display range
violation
< -1999 mV or > 1999 mV
pH display range violation
< -2 or > 16
mV range
Temperature range
violation
Sensocheck glass
Sensocheck ref.
Load error
Output current 1
< 3.8 mA
Output current 1
> 20.5 mA
Output current 2
< 3.8 mA
Output current 2
> 20.5 mA
Temperature value outside
table