Monitor horizontal levelling
1.
Choose a room ±10 m long
2.
Place the instrument next to wall no. 1.
3.
Switch on the instrument and let it self-level.
4.
Mark the horizontal line in the middle of the cross on wall no 1.
5.
Turn the instrument through 180° and mark the laser beam in the middle of the cross on wall no. 2.
6.
Move the instrument as close as possible to wall no. 2.
7.
Adjust the height of the laser so that the middle of the laser cross is on the wall no. 2 marker.
8.
Then turn the instrument through 180° and note the di erence between the middle of the laser cross and the marker on wall no. 1.
9.
This di erence should not exceed 2 mm.
10. NB: The tolerance depends on the distance of the walls between which the inspection has been carried out. This distance should be multiplied by 2.
Hence, if the instrument has an accuracy of 2 mm / 10 m, then in our example the calculation is: 10 m distance x 2 = 20 m. Tolerance is 4 mm / 20 m.