Description Of The Alarms - Cattani Maxi-Smart Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

• Description of the alarms

Alarm code
AC300
00
0
Microcontroller memory alarm
I14
2
95% amalgam level exceeded
I15
3
100% amalgam level exceeded
I00
32
Microcontroller memory alarm
I01
33
Short-circuit due to one of the two
motors
I02
34
Short-circuit before motor control
I03
35
Failed capacitor charging
I04
36
Temperature threshold exceeded
I05
37
Blower current limit exceeded
I07
39
Maximum voltage on capacitors
exceeded
S08
40
Centrifuge short-circuit
S09
41
Short-circuit of card in centrifuge
outlet
S10
42
Instantaneous current centrifuge
exceeded
S11
43
Delayed current limit of centrifuge
exceeded
I16
48
Vacuum sensor pipe disconnected
54
DESCRIPTION
SOLUTION
Contact technician
Replace amalgam container as soon as possible
Replace amalgam container
Contact technician
Check origin of short-circuit and eliminate it
Contact technician (card probably damaged)
Contact technician (replace card)
Ventilate plant room
Check effi ciency of blower (shutdown or excessive
resistance)
Check mains voltage max. 500V
Remove the short-circuit
Replace the card
Eliminate siphons in pipes or check effi ciency of
centrifuge (shutdown or excessive resistance)
Eliminate siphons in pipes or check effi ciency of
centrifuge (shutdown or excessive resistance)
Connect vacuum pipe to centrifuge

Publicité

Table des Matières
loading

Table des Matières