DC Voltage
Range
With batteries
7.0V to 999.9V
(1)
Without batteries
35V to 999.9V
(1) Measurement without batteries is only available for < 35°C,
> -15°C. The meter will show "bAtt" and ELV LED when
measurement is not available.
Max. Input Current : < 3.5mA @ 1000V
Overload Protection : AC/DC 1000V
AC Voltage
Range
(1)
With batteries
6.0V
to 999.9V
(2)
Without batteries
45V to 999.9V
(1) For > 65Hz, the minimum range is 8.0V.
(2) Measurement without batteries is only available for < 35°C,
> -15°C. The meter will show "bAtt" and ELV LED when
measurement is not available
Frequency Response : 45Hz to 400Hz
Max. Input Current : < 3.5mA @ 1000V
Overload Protection : AC/DC 1000V
Resistor
Range
Resolution
9999Ω
50.00kΩ
0.01kΩ
Output Voltage : about 0.5V
Overload Protection : AC/DC 1000V
Continuity
Continuity : The built-in buzzer sounds in the case of
measuring resistance less than 1.8kΩ and may be up to 2.7kΩ.
LED RX shows at the same time.
Continuity Indicator : 2.7kHz Tone Buzzer and RX LED
Response Time of Buzzer : < 100 msec.
Output Voltage : about 0.5V
Overload Protection : AC/DC 1000V
Resolution
Accuracy
0.1V
±(1.0% + 2D)
0.1V
Resolution
Accuracy
0.1V
±(1.5% + 5D)
0.1V
Accuracy
1Ω
±(1.5% + 5D)