An example of the correctly entered modification
1. When you have entered „–5", then each time the
device is turned on, the display will show the
following message „CAL" –5. This means that
the device is changing the measuring range from
10%-80% to 5%-75%.
Back to the factory settings/cancelling the
modification
Procedure:
1. Press and hold four buttons at the same time:
"MAX", "MIN", "T/M" and "ON/OFF".
2. Hold 4 buttons pressed for about 1 second.
Then release the buttons – letters "CAL" will
be displayed.
3. Using "MAX" and "MIN" buttons set the number
399 and press "T/M".
4. The display will show the previously entered
value of the modification, e.g. "+2". Enter "0" and
confirm the entered value with „T/M" button,
then turn the instrument off. The device returns
to factory settings, i.e. the measuring range will
be 10%–80%.