Télécharger Imprimer la page

DRAMINSKI HMMP Mode D'emploi page 13

Publicité

Les langues disponibles

Les langues disponibles

Modifying the moisture indications readings
This option is used for the modification (adjustment)
of moisture readings by increasing or decreasing
the displayed value (of the device) by the same
value in the entire measuring range. However, by
modifying the displayed values we are changing
the measuring range limits, i.e., when we make
an adjustment by „+2%" the measuring range will
change from 10%-80% to 12%-82%.
The modification is made, if the user finds that
for a given force of the pressed bale the tool has
a tendency to over- or understate the results by
the similar value in the entire moisture range. This
modification is useful when the bales of hay/straw
are poorly or very hard pressed, because the HMM
curve is based on an average compression strength.
12
Procedure:
1. Press and hold four buttons at the same time:
"MAX", "MIN", "T/M" and "ON/OFF".
2. Hold 4 buttons pressed for about 1 second.
Then release the buttons – letters "CAL" will
be displayed.
3. Using "MAX" and "MIN" buttons set the number
399 and press "T/M".
4. When the display shows „0" the device is ready
for calibration.
5. Enter the requested adjustment, keeping the
range from – 5 % to +5 % (e.g. „+2" or „–5")
6. Confirm the entered value with „T/M" button,
then turn off the device.

Publicité

Chapitres

loading