Télécharger Imprimer la page

Respironics MiniElite Mode D'emploi page 4

Publicité

Les langues disponibles

Les langues disponibles

Troubleshooting
Problems
Compressor overheats. The compressor
has an internal thermal protector to
prevent the device from overheating. If the
internal temperature reaches the preset
level, the unit will automatically shut down,
and the ON/OFF button light will turn red.
The compressor does not operate.
Medication does not nebulize properly.
Compressor does not work, and the
ON/OFF button light is red.
The device is noisy.
Should the device still not operate properly after checking the unit as indicated
above, contact Respironics Customer Service at 1-800-345-6443 or 1-724-387-4000.
Solutions
Turn the unit OFF for a minimum of 40
minutes to allow the compressor to cool.
Check all power sources to the compressor.
Push the ON/OFF button again.
Check for proper connection of the
tubing adapter.
Make sure the nebulizer is not empty.
Clean the nebulizer as described in
the manual.
Replace the nebulizer.
Check the intake filter and replace it if it
is dirty.
The thermal protector has been activated
due to:
a. The device operating outside the
conditions outlined in this manual.
b. The device has been operating near
heat sources or in hot environments.
Allow the compressor to cool down for
a minimum of 40 minutes. Wait until the
red light turns off before turning the
compressor on again.
Check that the filter cover is correctly
positioned.
Specifications:
Power Supply:
Weight:
Size:
Maximum pressure:
Maximum flow rate:
Mode of operation
Class II Device (Double Safety Insulation).
Type B Device (device with specific protection against electrical hazards).
Certification:
Electric Safety Standards EN 60601-1
Electromagnetic Compatibility according to EN 60601-1-2
Storage Conditions:
Temperature:
Humidity:
Operating Conditions:
Temperature:
Humidity:
Replacement Parts and Accessories:
Filter
1036898
AC power/battery charger
1036899
12V DC (Car) Adapter [1.25 A (min.)] 1036519
Rechargeable Lithium Ion Battery
RDD489
Symbol Glossary:
Attention: Read accompanying documents
~
Alternating Current
Optional Rechargeable Battery:
Operating time with battery
Up to 90 minutes
Number of treatments per charge
Up to 12 treatments when nebulizing 2.5 ml of medication
Normal time to charge
Approximately 6 hours
Electromagnetic Information
Portable and RF communication devices such as cellular phones, pagers, etc. can interrupt operation of electrical
medical equipment. For this reason, your compressor must be placed far enough away from these devices to prevent
interference. This device meets the IEC60601-1-2 Standard for Electromagnetic Compatibility (EMC). EMC data
sheets (part no. 1036799) are available upon request from Respironics, Inc. Customer Service at 1-800-345-6443 or
1-724-387-4000 or electronically at www.minielitecompressor.com.
100VAC - 120VAC~50/60 Hz
0.9 lbs.
4.5" x 3.8" x 1.9"
30 psi
8 LPM
Intermittent use 30 min ON/30 min OFF
MIN -10°F - MAX 158°F
MIN 10%RH - MAX 95%RH
MIN 50°F - MAX 105°F
MIN 10%RH - MAX 95%RH
ON/OFF (power)
SN
Serial Number

Publicité

loading