Do not use with an extension cord.
Keep the power cord away from heated surfaces.
Never block any opening of this product or place any objects into any
product opening.
This unit has been designed for intermittent use: 30 minutes ON/30 minutes OFF.
To avoid the risk of possible electrical shock, the nebulizer must be operated using
the specified power source.
Do not remove the compressor housing or disassemble unit. There are no
serviceable parts.
Never use while sleeping or drowsy.
Do not use while operating a vehicle.
Never drop or insert any object into any openings.
Do not use where oxygen is being administered in a closed area such as an
oxygen tent.
Use the nebulizer that is supplied with the compressor. Failure to do so may
compromise performance.
SAVE THESE INSTRUCTIONS
How To Use Your MiniElite
1. Check the medication nebulizer cup for damage. (The nebulizer enclosed with your
MiniElite should be replaced regularly. Ask your healthcare provider for replacement
guidelines and supplies.)
2. Place your MiniElite compressor on a firm, flat surface near an electrical outlet (if
using AC power) and where you can be seated comfortably during your treatment.
3. To use the compressor with the AC power adapter, connect the AC power adapter
into the power plug-in on the back of the compressor. Then connect the other end
into the electrical outlet.
4. Alternatively the unit can be used with the (OPTIONAL) rechargeable battery pack
or 12V DC (car) adapter. Read the instruction guide included with the rechargeable
battery pack or 12V DC (car) adapter carefully.
5. Do not operate the unit without the filter or with a clogged (dirty) filter. Operating
the unit with a dirty filter can impair proper operation and cause damage that is
not covered by the warranty.
6. Wash your hands before beginning your treatment.
7. Assemble your nebulizer and add your prescribed medicine following your
prescription and the separate nebulizer instructions.
8. Connect one end of the tubing to the bottom of the nebulizer and the other end
of the tubing to the air outlet on the compressor.
9. Turn the power switch on O to begin your treatment.
10. When your treatment is complete, turn the power switch off O and unplug the unit if
using AC power or 12V DC (car) adapter.
11. Clean the nebulizer according to its instructions.
Care and Cleaning of Your MiniElite
CAUTION: To reduce risk of electric shock, DO NOT DISASSEMBLE and never
submerge or place under running water.
1. Unplug the compressor before cleaning.
2. Use a damp cloth to wipe down the outside of the compressor at least
once each month (or more if necessary). Do not use abrasive cleaners.
3. Check the filter at least once each week. Replace the filter once a month or when
dirty.
4. To change the filter:
• Use a coin to unscrew the filter holder (Note: Turn CLOCKWISE to unscrew).
• Pull out filter and replace with new filter.
• Screw filter holder back on by turning COUNTER CLOCKWISE.
5. After each cleaning, check that the unit is working properly by verifying that
air still comes out of the nozzle when the unit is turned to the "on" position and
that the unit's noise level has not increased. (Increased noise indicates improper
filter positioning.)
6. The compressor is oil-free. Do not lubricate.
MiniElite Compressor
(back view)