Following machine cleaning and before sterilization, each part must be visually inspected for
cleanness. Any residues found must be removed manually.
Manual cleaning in an ultrasonic bath is also possible. However, machine cleaning should
always be preferred to manual cleaning.
NOTICE
Following machine cleaning and prior to sterilization, the various parts must be sprayed with
medical white oil (e.g. using KaVo spray).
32
The gear unit and gear cap (7, 8) must be
sprayed from the outside.
The handle (3) must be sprayed from the
inside.
Insert the INTRA coupling adapter of the
spray nozzle into the motor connecting
socket.
Flush the handle until clean oil is coming out
at the other end.
Instructions for Use
Angled Screwdriver ANGULUS
V. 2.0