Télécharger Imprimer la page

Tecno-gaz Multisteril Mode D'emploi page 45

Masquer les pouces Voir aussi pour Multisteril:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ENGLISH
Use of disinfectant / detergent other than that contained into the reservoir:
Disinfectant / detergent other than that contained in the reservoir can be used.
In the menu of the cycle to be performed enter 'NO' as reference to the detergent
automatic infeed.
Start the cycle, after some minutes (at the end of the water drain phase) the device sends
out an acoustic signal, it is now possible to open the lid and inject into the tank the correct
detergent percentage (1% - 60ml).
Close the lid to complete the cycle.
Disinfectant / detergent reservoir draining:
During the appliance use, when injection of detergent / disinfectant is required, in order to
drain completely the detergent / disinfectant reservoir, press and hold down button
SELECT
and at the same time hold down button OK.
After 5 seconds the peristaltic pump starts, during this phase an intermittent acoustic
signal is sent out which stops only when the phase is over, the detergent is drained directly
into the tank through the nozzle on the lid (I).
After about 10 min slightly open the lid (without letting the lid opening sensor intervene)
and make sure no detergent comes out of the nozzle.
Once the draining is completed, stop the process using button STOP.
17

Publicité

loading