10.2 Errors

In case of error alarm the device has to be reset to start a new cycle.
Display
22

ALARM RESET

Causes
Water infeed anomaly
-) cock closed
-) filters obstructed
-) water system pressure
anomaly
Water level anomaly
-) the device has been
moved or tilted during the
cleaning.
-) many tools have been
added after water infeed.
Water drain anomaly
-) filters obstructed.
-) drain hose in a wrong
position.
-) Wrong installation: anti-
return valve on the drain.
Hold down for 8s. buttom OK
Solutions
-) Make sure the cock is open.
-) Clean tank filter, drain filter, and
water infeed filter.
-) Make sure water pressure ranges
within
indicated values. In case of
insufficient
pressure
plumber.
-) Make sure the Water Block is not
locked.
-) Make sure there is no water into the
tank and start a new cycle
-) Make sure there is no water into the
tank and start a new cycle.
-) Do not add tools during the cycle.
-) Clean the tank filter and the drain
filter.
-) Position the drain hose correctly, see
installation.
-) Make sure there is no water into the
tank and start a new cycle.
apply
to
a