12. Specifications
Specifications
HF sensor, hydrogen fluoride (option)
Measuring principle
Measuring range
EN
Accuracy
Resolution
Permissible humidity
Max. zero offset
Long-term stability
Service life
H₂S sensor (option)
Measuring principle
Measuring range
Accuracy
Resolution
Permissible humidity
Max. zero offset
Long-term stability
Service life
CO sensor (option)
Measuring principle
Measuring range
Accuracy
Resolution
Permissible humidity
Max. zero offset
Long-term stability
Service life
Precision pressure sensor (option)
Measuring range
Accuracy
Non-linearity
(per IEC 61298-2)
38
SF₆-Q-Analyser
Electrochemical
0 ... 10 ppm
v
±1 ppm
v
0.1 ppm
v
15 ... 90 % r. h. (non-condensing)
0.1 ppm
v
< 2 % signal degradation/month (linear)
2 years starting from installation
Electrochemical
0 ... 100 ppm
v
±5 ppm
v
0.1 ppm
v
15 ... 90 % r. h. (non-condensing)
0.1 ppm
v
< 2 % signal degradation/month (linear)
2 years starting from installation
Electrochemical
0 ... 500 ppm
v
±9 ppm
v
0.1 ppm
v
15 ... 90 % r. h. (non-condensing)
0.1 ppm
v
< 2 % signal degradation/month (linear)
2 years starting from installation
0 ... 10 bar abs.
≤ ±0,05 % of span
Including non-linearity, hysteresis,
non-repeatability, zero offset and
end value deviation (corresponds to
measured error per IEC 61298-2).
Calibrated in vertical mounting position
with process connection facing
downwards.
≤ ±0,04 % of span BFSL
WIKA operating instructions analytic instrument, model GA11
g³-Q-Analyser
N/A
N/A
N/A
N/A