Table des Matières

Publicité

will sound on all switch positions except 'OFF' and
inhibited except for Insulation tests, which will remain available until the voltage
exceeds 55 V. (BM224 will beep and display a flashing 'V'. Testing is inhibited).
Default Voltmeter Ranges: 25V - 450V a.c. 50/60 Hz ± 2% ±3 digits
Automatic Discharge: Capacitive circuits are automatically discharged
when the test button is released following an insulation test.
Power Supply:
Battery life:
Battery cells should not be left in an instrument which may remain unused for
extended periods of time.
Low Battery Indicator: The low battery indicator symbol
Battery Replacement:
leads are connected. To remove the rear cover, release the screw at the
bottom of the cover and lift the cover upwards.To avoid the possibility of
shock, do not press the test button or touch the fuse when changing
batteries.
Fuse:
To check this fuse, select MΩ, open circuit the test leads and press the test
button until a reading is obtained. Display of the fuse symbol
code indicates a ruptured fuse. Located behind the rear cover, this fuse can
be replaced by the user. The rear cover must not be opened if test leads are
connected. The replacement fuse must be of the correct type and rating. To
avoid the possibility of shock, disconnect the battery before touching the fuse.
Safety:
The instruments meet the requirements for double insulation to IEC
1010-1 (1995) EN 61010-1 (1995) to Category III*, 300 Volts phase to earth,
440 Volts phase to phase, without the need for separately fused test leads.
*Relates to transient overvoltage likely to be found in fixed installation wiring.
E.M.C.
Weight:
Dimensions:
Cleaning:
Supplied
User Guide
Synthetic zip-up pouch
Test leads, prods & crocodile clips
Optional
Fuses (500mA) x 5
Fused probe kit FPK8
Publications Available
'A Stitch in Time'
450V - 600V a.c. 50/60 Hz ± 3%
d.c. - unspecified
6 x 1,5V cells IEC LR6 type only.
Typically 3000 x 5 second operations (worst
case, with
not selected).
when the battery cells are exhausted.
The rear cover must not be opened if the test
500 mA (F) H.B.C.10 kA min (32mm x 6mm)
In accordance with IEC61326 including amendment
No. 1
530g (including batteries)
195mm x 98mm x 40mm.
Wipe disconnected instrument with a clean cloth
dampened with soapy water or Isopropyl Alcohol
(IPA).

ACCESSORIES

7
. All selected tests will be
will appear
or an error
Part Number
6172-231
6420-132
6220-437
6121-289
6111-218
AVTM21-P8B

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Bm221Bm222Bm223Bm224Bm227

Table des Matières