Command: STOP
Source: PC
Destination: Controller
02
80
30
30
30
STX ADDR
WINDOW
Source: Controller
Destination: PC
02
80
06
03
38
STX ADDR ACK ETX
Command: SOFT-START (ON)
Source: PC
Destination: Controller
02
80
31
30
30
STX ADDR
WINDOW
Source: Controller
Destination: PC
02
80
06
03
38
STX ADDR ACK ETX
Command: SOFT-START (OFF)
Source: PC
Destination: Controller
02
80
31
30
30
STX ADDR
WINDOW
Source: Controller
Destination: PC
02
80
06
03
38
STX ADDR ACK ETX
31
30
03
42
WR
OFF ETX
CRC
35
CRC
31
31
03
42
WR
ON
ETX
CRC
35
CRC
31
30
03
42
WR
OFF ETX
CRC
35
CRC
Window Meanings
N.
Read/
Write
000
R/W
32
008
R/W
100
R/W
101
R/W
102
R/W
32
103
R/W
104
R/W
105
R/W
106
R/W
107
R/W
33
108
R/W
109
W
110
R/W
111
R/W
120
R/W
79/90
TECHNICAL INFORMATION
Data
Description
Type
L
Start/Stop
Start = 1
(in remote mode
Stop = 0
the window is a
read only)
L
Remote
Remote = 1
(default) or Serial
Serial = 0
configuration
(default = 1)
L
Soft Start
YES = 1
(write only in
NO = 0
Stop condition)
N
Set Point type
0 = Frequency
1 = Current
2 = Time
(default = 0)
N
Set Point
(default = 867)
threshold
(expressed in Hz,
mA or s)
N
Set Point delay:
0 to 99999
time between the
(default = 0)
pump start and
the set point
check (seconds)
L
Set Point signal
0 = high level ac-
activation type:
tive
the signal can be
1 = low level ac-
"high level active"
tive
or "low level ac-
(default = 0)
tive"
N
Set point
0 to 100
hysteresis
(default = 2)
(in % of threshold)
L
Water cooling
0 = NO
1 = YES
L
Active Stop
0 = NO
(write only in
1 = YES
stop)
N
Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
L
Pump life/ cycle
To reset write '1'
time/ cycle num-
ber reset
L
Interlock type
Impulse = 0
(default = 1)
Continuous = 1
L
Analog output
0 = frequency
type: output volt-
1 = power
age signal pro-
(default = 0)
portional to
frequency or
power
N
Rotational fre-
150 to 963
quency setting
(default = 963)
(Hz)
87-900-946-01(G)
Admitted
Values