5. SPECIFICATIONS

5.1
Electrical
Infrared
Temperature Scale: Celsius (°C) or Fahrenheit (°F) user-selectable
G
Temperature Range: -4°F to 1022°F (-20°C to 550°C)
G
Display Resolution: 0.1°C/°F or 1°C/°F
G
Accuracy: ±2% of reading or ±6°F/3°C, whichever is greater at 64.4
G
to 82.4°F (at 18 to 28°C) ambient operating temperature.
Temperature Coefficient: Changes in accuracy operating temperature
G
above 82.4°F/28°C or below 64.4°F/18°C: ±0.2% of reading or ±0.36°F/
0.2°C, whichever is greater.
Response Time: 1 second
G
Laser: Red, <1mW (670nm) Class II, 2 - 50ft Range.
G
Spectral Response: 6 to 14µm nominal
G
Emissivity: Pre-set at 0.95, user selectable from 0.10 to 1.00
G
Detection Element: Thermopile
G
Optical Lens: Fresnel Lens
G
Field of View (FOV) ratio: 10:1 (Distance: Diameter)
G
K Type
Temperature Scale: Celsius (°C) or Fahrenheit (°F) - User-selectable
G
G
Temperature range of the instrument in function of the K type
thermocouple chosen (see § 6) : - 40 to +1350°C (-40 to 2000°F)
-40 to 200°C (-40 to 392°F) with the K type thermocouple supplied
as standard.
G
Resolution: 0.1°C/°F or 1°C/°F
G
Accuracy (meter):
-40°C to 1350°C: ±(0.1% rdg ± 1°C) on -40°F
to 2000°F: ±(0.1% rdg ± 2°F) (Accuracy is specified for operating
temperatures over the range of 18°C to 28°C, for 1year, excluding
the sensor)
Temperature Coefficient: 0.1 times the applicable accuracy
G
specification per °C from 0°C to 18°C and 28°C to 50°C
Input Protection: 24VDC or 24Vrms maximum input voltage on any
G
combination of inputs.
Sample Rate: 2.5 times per second.
G
Input Connector: Standard miniature thermocouple connectors (flat
G
blades spaced 7.9mm, center to center).
Temperature Response: Temperature indication follows Reference
G
Temperature/Voltage Tables N.I.S.T. Monograph 175 Revised to ITS
-90 for K-type thermocouples.
5.2 General specifications
G
Display: 3½ digit liquid crystal display (LCD) with maximum reading
of 1999
G
Over-Range: (OL) or (-OL) is displayed
Low battery indication: " symbole batterie " is displayed when the
G
battery voltage drops below the operating level
G
Sample rate: 2.5 times per second, nominal.
Environment:
G
Operating:
32°F to 122°F (0°C to 50°C) at < 80% relative humidity
Storage:
-4°F to 140°F (-20°C to 60°C), 0 to 80% RH with
battery removed from meter
Auto power OFF: 15 seconds approx.
G
Altitude: 2000m max.
G
G
Input Protection: 24VDC or 24rms maximum input voltage on any
combination of inputs.
G
Battery: Standard 9V battery (NEDA 1604, IEC 6F22 006P , or equivalent)
Battery Life: 100 hours (continuity) typical with carbon-zinc battery
G
(Back-Light not illuminated)