Troubleshooting

Motor not running
Check if:
– the Vario 18 is switched on.
– the mains voltage is correct and if the mains plug is
inserted correctly into the fixed mains socket.
– the internal battery is charged (only AC / DC version
under battery operation)
Suction of pump too low
Check if:
– the vacuum regulator is set correctly.
– the tubings are not defective or broken. If necessary,
replace.
– all plug-in connections are tight.
– the floater on the reusable lid moves freely up / down.
– the suction jar / lid has no cracks, brittle areas,
discolouration. Replace if necessary.
– the disposable system has no cracks, brittle areas,
discolouration. Replace if necessary.
– filter is not clogged. To test if the filter is clogged see
chapter «filter test».