EN-8
Sample Pans (50/Box)
Pads, Glass Fiber (200/Box)
Reusable Sample Pans (3/pk)
Reusable Deep Sample Pans (3/pk)
Reusable Wire Cage
Sample Pan Handler
50g Calibration Weight
6. TECHNICAL DATA
Admissible Ambient Conditions
Location:
Temperature:
Relative humidity:
Warm up time:
Height above sea level: Up to 2000 m
Power Input:
Voltage fluctuations:
Power load:
Power line fuse:
Protected against dust and water, Pollution degree: 2, Installation category: Class II
Specifications
MODEL
Capacity
Readability (% requires >10 g sample)
Temperature Settings
Heat Source
Calibration - Weight
Calibration - Temperature
Pan Size
Unit Dimensions (DxWxH)
Unit Weight
Shipping Dimensions (LxWxH)
Shipping Weight
Communication
RS232 Pin Connections
GND
5
9
RS232 Data Settings
Baud Rate (default ): 2400 Data Bits: 7
Baud Rate:
RS232 Commands
The RS232 Interface allows a computer to control the Analyzer, as well as to receive data such as displayed weight.
80850086
80850087
80850088
80252479
80252477
80252476
51054-16
Indoor use only
10° C to 40° C
15 % to 80 % at 30° C non-condensing
At least 15 minutes after connecting the Analyzer to the AC supply;
when switched from standby mode the Analyzer is ready for immediate use.
100 VAC – 120 VAC, 3 A, 50/60 Hz or 200 VAC -240 VAC, 3 A, 50/60 Hz
-15% +10%
MB25: 250 W; MB23:300 W (Maximum during drying process)
1 piece, 5 x 20 mm, 2.5 A 250 V
0.01 g, 0.1% (100-110 g)
Female DB9 connector
RxD
TxD
Pin 2: Analyzer transmit line (TxD)
1
Pin 3: Analyzer receive line (RxD)
6
Pin 4 and 6: Internally connected
Pin 5: Ground signal (GND)
CTS
Pin 7: Clear to send (hardware handshake) (CTS)
RTS
Pin 8: Request to send (hardware handshake) (RTS)
Parity: N
9600 Data Bits: 8
Parity: N
CBM910 Impact Printer, JP
CBM910 Impact Printer, US
CBM910 Impact Printer, EU
Cable, STP103
Cable, CBM910
Data collection Software
MB25
0.005 g, 0.05% (0-99 g)
50° C to 160° C (5° increments)
Halogen
OHAUS Temperature Calibration Kit
90 mm diameter
11.0x6.5x5.0 in / 28.0x16.5x12.7 cm
4.6 lb / 2.1 kg
20.2x15.5x13.5 in / 51x40x35 cm
10 lb / 4.5 kg
Stop Bits: 2
Stop Bits: 1
MB23 / MB25
80252041
80252042
80252043
80252581
80252571
SW12
MB23
110 g
0.01 g, 0.1%
Infrared
50 g
Flow Control: Xon/Xoff
Flow Control: Xon/Xoff

Publicité