Control box + Battery
Producer
Voltage
Capacity
Encapsulation
Colour
Electrical safety
IEC60601-1 3rd revision
20. Hand Control Plus
**1. Service indicator – flashes red after 11 months' operation.
Constant light (red) after 12 months' operation.
The indicator is a reminder to review the maintenance instruction (see Maintenance
Instructions). This should be done at least once every 12 months.
Reset: hold down buttons <2a> + <2b> for at least 30 seconds.
**The indicator will go out around 30 seconds after the hand control is
activated (battery conservation function).
21. Expected lifetime:
The life expectancy of the product is 8 years.
I F U
TiMOTION
24V
2.0AH
IP66
Black
Class II
2a
2b
1
13