Electromagnetic Compatibility Information
Table 1
Guidance and manufacture's declaration – electromagnetic emissions
The SBM 21 is intended for use in the electromagnetic environment specified below.
The customer or the user of the SBM 21 should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The SBM 21 is intended for use in the electromagnetic environment specified below. The customer or the user of the SBM 21 should assure that
it is used in such an environment.
IMMUNITY test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Electromagnetic environment – guidance
Group 1
The SBM 21 uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equip-
ment.
Class B
The SBM 21 is suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buil-
Not applicable
dings used for domestic purposes.
Not applicable
IEC 60601 test level Compliance level Electromagnetic environment – guidance
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30 %.
Power frequency magnetic fields should be at levels characteristic of
a typical location in a typical commercial or hospital environment.