6 Function check
6
Function check
6.1 Intervals
6.2 Perform function check
Requirement
Result
30
EN
SOMNOalarm
Perform a function check at regular intervals:
•
before every use
•
after each hygiene treatment
Device has been disconnected from the therapy device.
1. Check device for external damage.
2. Put plug in serial port of therapy device.
An automatic self-test runs which simultaneously covers the
points below:
•
a sequence of beeps sounds
•
the alarm indicator is green
3. If one or more conditions are not met: send the device to your
specialist dealer or to Weinmann.
Function check is complete.