6 Function check
6
Function check
6.1 Intervals
6.2 Performing a function check
Result
34
EN
prismaPSG
Carry out a function check at regular intervals:
•
After each hygienic preparation
•
After each repair
•
At least every 6 months
1. Check whether all the modules are set up and connected
correctly in accordance with the instructions for use.
2. Check the housing for cracks, damage, and severe soiling.
3. Connect the module up to the therapy device
Connecting up the prismaPSG module", page
4. Check whether the green
therapy device display.
5. Check whether a value corresponding to the information on
the therapy device display is shown on the pressure channel
being fed-in.
6. If the module does not work properly or shows signs of
damage: Contact your authorized dealer.
The function check is performed.