5
Hygiene treatment
5.1 General information
5.2 Intervals
5.3 Hygiene treatment for device
Part
Device
Result
•
Use suitable gloves (e.g. household or disposable gloves) when
disinfecting.
•
Follow the instructions for use for the disinfectant used.
•
Recommended: terralin
Clean the device in accordance with the specifications for the
therapy device connected.
Risk of material damage as a result of ingress of liquids!
Liquids which penetrate the device may damage it.
Do not immerse the device in liquids.
Protect the plug from moisture.
1. Disconnect the plug from the serial port of the therapy device.
2. Subject the device to a hygiene treatment in accordance with
the table below.
Cleaning
Wipe down: use water or
mild detergent.
3. Put the plug in the serial port of the therapy device
Connect device",
4. Perform a function check
The device has been subjected to hygiene treatment.
®
protect for disinfecting by wiping.
Disinfecting
Disinfect by wiping on
change of patient
page 27).
(see "6 Function check",
SOMNOalarm
5 Hygiene treatment
Sterilizing
Not permitted
(see "4.1
page 30).
EN
29