As you continue to pull the hand lever slowly, the tractive force increases, which works upon the cable as a
During the measurement, the force progression can be read off
the display.
The peak value, which mostly appears before the connection
snaps up, is captured at 1000Hz and is indicated when the PEAK-
button is pressed (so long as the drag indicator function was not
active).
Depending on the wire strength and the related standard, the following tensile strength should at least be
achieved.
AWG
Cross-Section
30
28
26
24
22
20
18
16
14
12
10
DIN 41611/3 is replaced by DIN IEC 60352 Part 2
MIL-T-7928 is replaced by SAE AS7928 Table II
BS5B178 corresponds to IEC 60352 Part 2
UL486A corresponds to IEC 60352 Part 2
Cable Diameter
0,06 mm²
0,36 mm
0,09 mm²
0,38 mm
0,14 mm²
0,48 mm
0,22 mm²
0,61 mm
0,34 mm²
0,76 mm
0,56 mm²
0,97 mm
0,93 mm²
1,27 mm
1,25 mm²
1,44 mm
1,93 mm²
1,80 mm
3,16 mm²
2,29 mm
4,65 mm²
3,10 mm
retention force. An additional tractive force develops, which works
upon the cable connection.
Increase the force until the cable snaps off at its most fragile
point, which should be the connection between the cable and the
test component.
SAE AS7928
Table II
32 N
45 N
67 N
85 N
170 N
223 N
312 N
490 N
18
IEC 60352 Part 2
6 N
11 N
18 N
28 N
40 N
60 N
90 N
135 N
200 N
275 N
355 N
UL 486 C
6 N
11 N
18 N
28 N
40 N
45 N
45 N
68 N
100 N
138 N