CMD600
6.2.2 Connect the system input/output (View of pins in CMD600)
1
2
8
7
3
4
5 6
The charge amplifier is supplied via this socket with an external DC voltage
source. The CMD600 does not have an On/Off switch. If the correct supply
voltage is connected, the LED lights up (see Page 12).
6.2.3 Ethernet connection
3
4
2
1
Category 5 shielded cables are used for the Ethernet connection. We
recommend the HBM Ethernet Cross Cable; Order number 1−KAB284−2
We recommend using an Ethernet Cross cable when the CMD600 is
connected directly to a PC or host.
A2781−2.1 en/de/fr
Pin No. Signal name
1
Ground supply
2
Sensorteach
3
Reset
4
TEDS
5
Charge out
6
Output ground
7
Not in use
8
Voltage supply
Pin No.
1
2
3
4
Description
−
Digital input, active High
Digital input, active High
−
Output signal
Output signal ground
Not in use
Voltage supply between Pin 8 and 1
Signal name
TX +
RX +
TX −
RX −
29
Value
Color
code
KAB168...
−
wh
+18 − 30 V
br
+18 − 30 V
gn
−
ye
"10 V
gy
−
pk
−
bu
+18 − 30 V
rd