3.1 System input/output
The charge amplifier is supplied with voltage (18−30 V DC) via this socket
(see Chapter 6.2.2 ) and the measurement signal is output with max. "
10VDC.
The charge amplifier is fitted with TEDS (Transducer Electronic Data Sheet).
In order to transfer TEDS information to the downstream evaluating
electronics, the TEDS chip must first be written to (see Chapter 3.7).
RESET/MEASURE
The measurement can be triggered with the RESET/MEASURE input and the
input capacitors of the charge amplifier cleared with (RESET) (see chapter
11.3 "Information about the charge amplifier and electrical connection").
SENSOR TEACH
SensorTeach is an additional function for automatic scaling of the charge
amplifier (see Chapter 3.6).
3.2 System LED
The LED shows the actual operating state of the charge amplifier.
CMD600 state
IP address not configured
Connection via Ethernet
Measure
Reset
Overload
SensorTeach function in
600,000 pC range
SensorTeach function in
6000 pC range
Ready for firmware update
Bootloader mode
Telnet connection
LED display
Green-blue,
Device IP address of factory setting is
flashing
Constant blue
Charge amplifier ready for measurement
and connected to a host system via Telnet.
Constant green
Charge amplifier ready for measurement,
not connected to a host system via Telnet.
Constant red
Charge amplifier in RESET mode (not ready
Red (green or
Overload error (output voltage greater than
blue)
flashing
Yellow,
SensorTeach activated, CMD600 in
flashing, 1 Hz
Yellow,
SensorTeach activated, CMD600 in 6000 pC
flashing, 2 Hz
White, flashing,
Bootloader is loaded. Factory settings can
2 Hz
be loaded within the next 3 seconds if the
Reset input is briefly actuated once.
Red, flashing,
CMD600 waiting for PC input to start the
1 Hz
Blue/Yellow/
Red/Green,
currently connected to Telnet or CMD Assi-
flashing
2 Hz
Note
loaded.
for measurement).
10 V).
600,000 pC range.
range.
firmware update, Timeout 10 sec.
Identification of the CMD600 that is
stent.
CMD600
A2781−2.1 en/de/fr