2. Safety
GB
Special conditions for safe use.
10
WARNING!
Residual media in dismounted pressure instruments can result in a risk to
persons, the environment and the equipment. Take sufficient precautionary
measures.
WARNING!
If the CPG1000 has been exposed to overload or sudden physical shock
(e.g. being dropped) it should be checked for signs of damage that may
cause concerns about safety. If there is any doubt, return the unit to WIKA
for evaluation. For information on returning the instrument see chapter
"10.2 Return".
The CPG1000 is not intended for use with flammable substances and must
be used only in locations which provide sufficient protection against the
intrusion of solid foreign objects or water which could impair the safety.
CAUTION!
To avoid possible damage to the calibration reference or to equipment
under test:
If the message "OL" is displayed, the measuring range has been
exceeded and the pressure source must be removed immediately
from the CPG1000 in order to prevent damage to the internal pressure
transducer.
The maximum permitted torque is 13.5 Nm = 10 ftlbs. This permitted
torque must NEVER be exceeded.
WIKA Operating Instructions Model CPG1000