Troubleshooting; Dépannage / Diagnóstico De Problemas / 故障排除 / Устранение Неполадок; Solução De Problemas / Diagnostica / Fehlerbehebung - Ralston Instruments SP0V Manuel D'utilisation

Pompe d'échantillonnage de gaz
Table des Matières

Publicité

Troubleshooting

Dépannage ......................................20
Diagnóstico de problemas ...............21
故障排除 ..........................................22
Устранение неполадок ..................23
The pump will pump up, but pressure slowly decreases
EN
If the pump will pump up, but pressure slowly decreases, then there is an
external leak. Follow these instructions to locate and repair the leak:
1. Connect the pump to a Device Under Test (DUT) with a Ralston
Quick-test™ hose.
2. Make sure the process connections are assembled wrench-tight.
3. Pressurize the Hand Pump.
4. Spray soapy water or leak detection fluid where leaks are suspected
or immerse the pump in water. Be careful not to immerse the pressure
gauge or calibrator.
5. Observe where bubbles are forming to determine where there is a leak.
6. Remove the leaking part and remove the cut or damaged O-ring.
7. Clean and lubricate the O-ring.
8. Replace the O-ring and reassemble.
Pumping Piston pushes out by itself and pressure
decreases
If the Pumping Piston pushes out by itself and pressure decreases, then
the internal Check Valve is not functioning properly. Follow the instruc-
tions to replace the internal Check Valve:
1. Remove the Pumping Piston.
2. Remove the Internal Check Valve.
3. Clean and lubricate the Check Valve.
4. Reinstall the Check Valve.
5. Reinstall the Pumping Piston.
When the Pumping Piston is pumped, pressure does
not increase
If when the Pumping Piston is pumped, pressure does not increase,
then the External check valve in the Inlet fitting may be leaking. Follow
these instructions to repair or replace the Inlet fitting:
1. Remove any hoses from inlet and outlet fittings.
2. Cap outlet fittings.
3. Attempt to pressurize the Hand Pump.
4. Spray soapy water or leak detection fluid on the inlet fitting or immerse
the pump in water.
Solução de problemas .....................25
Diagnostica ......................................26
Fehlerbehebung ..............................27
19

Publicité

Table des Matières
loading