Fastening of light therapy units LD 72 and LD 110 on the wall
The light therapy units LD 72 and LD 110 can be mounted on the wall.
On the back side of the top of these light therapy units there is an opening in the middle below the air vents (see illustration).
The base of the light therapy units LD 72 and LD 110 must be unscrewed before the fastening. Therefore the right and the
left thumb screw must be completely removed from the foot. So the foot can be removed completely from the upper part of
the unit. Afterwards the screws have to be screwed in for safekeeping.
Operation of the light therapy units with the DAVITON
®
The activated DAVITON
(not included in the scope of delivery) reminds you with an acoustic signal every 90 seconds to
look for 3 seconds direct into the light source.
Frequency and time of treatment
If the treatment takes place at your practitioner it should initially be carried out 5 times per week for a period of 2 – 3 weeks;
thereafter, 3 times per week for a period of 2 – 3 weeks. After that, the treatment should take place as needed or when the
symptoms occur again. At home, the treatment can take place daily, especially during autumn and winter time. The best
treatment time for winter depression is the early morning or the time of the lowest mood.
Side effects and contraindications
By using phototherapy the following side effects have been observed: Headaches, nausea, irritability and burning of the
eyes can occur rarely for a short period. An absolute contraindication that means circumstances that do not allow the use
of light therapy units is not known so far.
Please take care in case of all types of eye diseases and the eye specialist should be consulted before. In case of patients
being under treatment with lithium or medicine for example Fluoxetine, Imipramine et al., which increase the sensibility of
the eyes on light a supervision of the light therapy by an eye specialist could be considered.
Safety Check
Only skilled and qualified companies are entrusted to perform safety checks. Suitable test devices and facilities
are mandatory. The medical device should only be used with accessories listed in the user manual.
In accordance with MPBetreibV, the light therapy boxes for use in the commercial sector has to undergo
mandatory safety checks every 2 years or in shorter intervals and after each repair or refurbishment. The operator
is responsible to repair (commission repair) of any defects detected during the safety check.
Cleaning, maintenance and recycling
The instruments do not need any special maintenance. To clean the outer casing parts of the devices, we recommend a
mild cleaning agent (as used for glass). First please disconnect the power supply.
A special maintenance is not necessary for these devices. If lamps must be replaced or repairs have to be done only
®
DAVITA
or a company which is qualified to do such works.
At this opening the units can be fixed on the wall with a screw.
Depending on the type of wall, an appropriate screw must be used,
may be with pegs. E.g. a round head wood screw which is screwed
into an expansion anchor. The screw and the anchor are not in-
cluded in the scope of delivery. The screw head must not exceed
a max. height of 6 mm and a max. diameter of 8 mm. The distance
from the wall to the screw head can be max. 8 mm. The screw shaft
may have a maximum diameter of 4 mm. The load on the screw
and the condition of the wall, to which the unit has to be mounted
must have a capacity of at least 3 kg.
- 8 -