Standard presetting:
Maximum stroke volume
→
Set the flow rate by turning the adjusting screw.
→
Fix the set flow rate with locking screw.
8.3.
functions
8.3.1.
control diagram
V1
V2
V3
0
10
15
ms
T0
T1
T2
Fig. 6:
Control diagram
Via the parameters (T0 – T6), the times for the valve control are set.
By changing the interval times (T6 – T1), the frequency can be
changed.
Type 7616
Start-Up
15
15
5
T3
T4
T5
T6
times for maximum throughput:
While maintaining the times illustrated in the figure "Control
diagram" and an interval time of 0 ms, the maximum frequency
(15 Hz or 800 cycles/min) will be reached.
This corresponds to a throughput of approximately 4 ml/min with a
set maximum stroke volume.
recommended times for maximum accuracy:
For dosing with the highest accuracy (Tolerance and reproducibility)
≤ ± 2%), the following times are recommended:
ms
0
T0
T1
This corresponds to a throughput of approximately 1.15 ml/min with
a set maximum stroke volume.
0
T1
30
50
50
50
T2
T3
T4
english
50
x
T5
T6
T1
13