Balluff BTL7-A/C/E/G-M-A/B/Y/Z8-NEX-S32/KA Série Notice D'utilisation page 54

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

BTL7-A/C/E/G _ _ _ -M _ _ _ _ -A/B/CD/Y/Z(8)-NEX-S32/KA _ _
Micropulse Transducer - Rod Style
7
Calibration using teach-in
NOTICE!
Interference in function
Teach-in while the system is running may result in
malfunctions.
► Stop the system before performing teach-in.
Initial situation:
Transducer with magnet within measuring range
1. Activate buttons
► Hold down any button for at least 3 s.
► Release buttons.
► Within 1 s, hold down
at least 3 s.
⇒ Output indicates error value.
⇒ Buttons are activated.
If an error or an interruption occurs while
activating the buttons, allow a wait time of 12 s
before retrying.
2. Select teach-in
Œ
► Hold down
for at least 2 s.
⇒ Indication for "Teach-in" is displayed.
Œ
► Release
.
⇒ Current position value is displayed.
3. Set null point
► Bring magnet to the new null point.
Œ
► Hold down
for at least 2 s.
⇒ The new null point is set.
4. Set end point
► Bring magnet to the new end point.
► Hold down
for at least 2 s.
⇒ The new end point is set.
5. Exit teach-in and deactivate buttons
Œ
► Hold down
and
⇒ Output indicates error value.
Œ
► Briefly press
or
⇒ Buttons are deactivated.
⇒ Current position value is displayed.
18
english
Œ
and
simultaneously for
simultaneously for at least 6 s.
(< 1 s).
LED legend:
LED display Displayed values (example)
LED1 LED2 At 0...10 V
1
2
> 3 s
< 1 s
1
2
> 3 s
> 2 s
1
1
> 2 s
2
> 2 s
> 6 s
1
2
1
2
< 1 s
LED not on
LED flashing green
LED green
LED 1 and LED 2 flashing green-green in
LED red
alternation
Ignition protection
"nA" and "tb"
At 4...20 mA
5.39 V
9.15 mA
5.39 V
9.15 mA
10.50 V
3.60 mA
4.00 V
12.00 mA
5.39 V
9.15 mA
1.04 V
4.82 mA
0.00 V
4.00 mA
0.00 V
4.00 mA
9.89 V
19.13 mA
10.00 V
20.00 mA
10.00 V
20.00 mA
10.50 V
3.60 mA
10.00 V
20.00 mA

Publicité

Table des Matières
loading

Table des Matières