13. Maintenance and care

Check once a month that the eye shower is in working order. (Guidelines for Labora-
tories BGI/GUV-I 850-0)
As a preventative measure to reduce microbiological contamination, it is recom-
mended to replace the water in the fitting at shorter intervals by opening the valve.
Use suitable cleaning agents which are not harmful to the fitting in a proper manner
and rinse with water after use. High-pressure cleaners must not be used for cleaning.

14. Troubleshooting

Fault
Water does not flow
Water volume too low
Jet height too high
Jet height too low
If you cannot rectify a fault or if it is not in the list, please contact our customer
service department!

15. Spare parts

Description
1 Hand shower complete. . . . . . . . . 2000105863
2 Pressure reducer . . . . . . . . . . . . . 2000109525
3 Filter (packaging unit10x) . . . . . . . 2000109401
4 Dust cap with ball chain . . . . . . . . 2000105849
5 Spray head (packaging unit 4x) . . 2000104725
6 Non-return valve
(packaging unit 10x) . . . . . . . . 2000104744
7 Seal (packaging unit 10x) . . . . . . . 2000104668
8 Connection hose. . . . . . . . . . . . . . 2000105845
Maintenance set consisting of:
O-rings . . . . . . . . . . . . . . . . . . 2000105864
6
Cause
– Water supply interrupted
– Dirty filter
– Pressure reducer defective
– Dirty filter
– Jet height set incorrectly
– Flow pressure too high
– Supply pressure too low
– Pressure reducer set incorrectly
– Flow pressure too low
Order No.
Remedy
➯ Restore it
➯ Clean and replace
if necessary
➯ Replace it
➯ Clean it
➯ Set it
➯ Assembly with
pressure reducer
➯ Check it
➯ Set it
➯ Assembly without
pressure reducer