CAL 2: The offset for sensor 2 (relative humi-
dity) is set with CAL 2 (one-point calibration).
The offset turns the characteristic around the
lower calibration point (11 % r.H.). The
calibration point must be in the range from 30 %
to 95 % r.H. The offset is displayed in the
upper display. A maximum ± 10 % r.H. can be
set. CAL 2 can only be selected in
connection with the unit % r.H. and at a
prevailing humidity of at least 30 % r.H.
Selection can be made with
The factory settings are obtained by setting the offset to 0.0.
page 38
HUMIPORT 05 IR english
and
, confirmation with
.