E+E Elektronik HUMIPORT 05 IR Notice D'utilisation page 37

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Manual
One-point calibration of temperature and relative humidity
With CAL a one-point calibration can be carried out for the sensor
displays of temperature (CAL 1) and relative humidity (CAL 2).
All sensors are already calibrated at the factory and have an
appropriate factory calibration characteristic.
In the one-point calibration a global calibration curve shift is made
by specifying an offset which is effective over the whole
measuring range.
The offset to be entered is the value by which the calibration
curve is shifted.
CAL 1: The offset for sensor 1 (temperature)
is set with CAL 1 (one-point calibration).
The offset is shown in the lower display.
A maximum ± 10 °C or ± 10 °F can be set.
Selection can be made with
confirmation with
The factory settings are obtained by setting the offset to 0.0.
page 37
HUMIPORT 05 IR english
and
,
.

Publicité

Table des Matières
loading

Table des Matières