KERN EOB_F Mode D'emploi page 45

Balances à plateforme / balance stationnaire
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

 In weighing mode press the UNIT key and keep it pressed until „CAL" followed
by the weight value of the required adjustment weight in displayed. „LOAd" is
displayed alternatively.
 Carefully place the required adjustment weight (see tab. 1 „Technical data") in
the centre of the weighing plate
Press TARE key, adjustment is started.
 Wait until the stability display appears, after that the adjustment is finished. The
balance will automatically return to Weighing mode upon successful completion
of the adjustment and will display the weight of the adjustment weight.
 Take away adjustment weight.
An error message will be displayed in the event of an adjustment error or
incorrect adjustment weight. Remove the adjustment weight and repeat the
adjustment process.
Keep the adjustment close to the balance. Daily control of the weighing
exactness is recommended for quality-relevant applications.
EOS_F-BA-e-1611
(Example)
(Example)
(Example)
15

Publicité

Table des Matières
loading

Table des Matières