Liquisys M CCM223/253
Zero calibration
CALIBRATION
(only if A1 = 120)
zero
(CCS120)
no;
Enter value of ORP
buffer
CALIBRATION
Buffer value of last cal.
ORP mV
0 ... 1500 mV
(dep. on B1)
Enter calibration
temperature
CALIBRATION
pH
25.0 °C
(dep. on B1)
Enter DPD
calibration value
CALIBRATION
Cl 2 / ClO 2
Value of last cal.
C
MEAS. VALUE
DISPLAY
with
TEMPERATURE
DISPLAY
in °C
Select sensor
140; 141; 240; 241;
963; 120
SETUP 1
A
Select operating mode
Off;
pH; ORP mV
SETUP 2
B
Controller switch-off
by current input
Off;
CURRENT INPUT
Z
Select current output
Output1;
CURRENT OUTPUT
O
Endress+Hauser
Calibration status is
displayed
yes
o.k.
Exxx
C15
C14
Calibrating
Adopt if stable
±1 mV (> 10 s)
C31
C32
Enter pH value
of first buffer solution
Buffer value of last cal.
–10 ... +60 °C
pH 3.50 ... 9.50
C21
C22
Zero calibration
(only if A1 = 963)
no;
yes
C11
C12
Meas. value display
Temperature display in °F
1st error is displayed
(if present)
Display unit
ppm; ppb
mg/l;
A1
A2
pH compensation
Off;
Manu; Auto
B1
B2*
* field B2 not available for CCS240/241
Delay for controller
sw.-off current input
Input
0 s
0 ... 2000 s
Z1
Z2
Select meas. parameter
for 2nd current output
Output2
°C;
pH; ORPmV; Contr
O1
O2
Store calibration
Operating matrix version EP
result
yes;
no; new
C15
Zero point is
Calibration status is
displayed
displayed
–100 ... +100 mV
o.k.
Exxx
C33
Enter pH value
Calibrating
of second buffer solution
Adopt if stable
Buffer value of last cal.
pH ±0.05 (> 10 s)
pH 3.50 ... 9.50
C23
Calibration status is
Slope is
displayed
displayed
100%
min. 25% (3%)
max. 500%
o.k.
Exxx
C13
Meas. value display
Meas. value display
No temp. display
pH;
ORP (dep. on B1)
Other errors are
displayed
(up to max. 10 errors)
Controller switch-off
Delay for contr.
by sample stream
sw.-off by sample stream
Off;
INS
0 s
0 ... 2000 s
A3
Manual
Enter correct
pH compensation
process temperature
last comp. value
current meas. value
pH 4.00 ... 9.00
0 ... 50 °C
B3
Delay for controller
Switch-off limit
sw.-on current input
for current input
0 s
0 ... 2000 s
50%
0 ... 100%
Z3
Select table
options
read
edit
Tab = Table
O3(3)
Enter simulation
value
current value
0 ... 22.0 mA
sim = simulation
O3(2)
Select current range
4–20 mA;
0–20 mA
lin = linear
O3(1)
Enter calibration
results
yes;
no; new
C34
C35
Calibrating
Adopt if stable
pH ±0.05 (> 10 s)
38.00 ... 65.00 mV/pH
C24
C25
Store calibration
result
yes;
no; new
C14
C15
Meas. value display
Meas. value display
pH signal in mV
Sensor current in nA
Delay for contr.
sw.-on by sample stream
ext.Hold
0 s
0 ... 2000 s
A4
A5
Enter temperature
difference (offset)
current offset
–5.0 ... + 5.0 °C
B4
B5
Switch-off direction
Feedforward control
for current input
to PID controller
Low;
High
Z4
Z5
Enter number of
table pairs
1 ... No. of table pairs
1
1 ... 10
O331
O332
O321
Enter 0/4-mA value
Enter 20-mA value
0.00 mg/l;
0 ... 20 (5) mg/l
2.0 (0.5);
0.00 ppm
; 0 ... 20 ppm
2.0 ppm;
0 ppb;
0 ... 20000 ppb
2000 ppb;
pH 4.00;
pH 4 ... 9
pH 9.00;
0 mV;
0 ... 1500 mV
1000 mV;
0 °C;
0 ... 50 °C
50 °C;
O311
O312
Annexe
Slope is
displayed
59.16 mV/pH
C26
Select
digital input 1
Clean Trig.
A6
Off;
lin = linear
Z6
Select table
pairs
1
asign
O333
0 ... 20 (5) mg/l
0 ... 20 ppm
0 ... 20000 ppb
pH 4 ... 9
0 ... 1500 mV
0 ... 50 °C
O313
A0002637-FR