GUIDANCE AND MANUFACTURER'S DECLARATION
Electromagnetic emissions
The « KINETEC Performa » is intended for use in the electromagnetic environment specified below.
The customer or the user of the « KINETEC Performa » should ensure that it is used in such an environment.
Immunity test
Conducted RF interference
IEC 61000-4-6
Radiated RF interference
IEC 61000-4-3
NOTE 1
At 80 and 800 MHz, the highest frequency range is applicable.
NOTE 2
These directives cannot be applied in every situation. Electromagnetic propagation is affected by absorption and reflection
by structures, objects and people.
The field intensity of fixed emitters such as base stations for radio-telephones (cellular/cordless) and land mobile radios, amateur
a
radio, AM/FM radio broadcasts and TV broadcasts cannot be predicted exactly in theory. To evaluate the electromagnetic
environment due to fixed RF emitters, an on-site electromagnetic investigation should be considered. If the field intensity measured
where the "KINETEC PERFORMA" splint is used exceeds the aforementioned applicable RF compliance level, the "KINETEC
PERFORMA" splint should be monitored to check that it is working normally. If abnormal results are observed, additional measures
may be necessary, such as reorienting or repositioning the "KINETEC PERFORMA" splint.
Over the frequency range from 150 kHz to 80MHz, field intensities should be less than 3V/m.
b
Recommended separation distances between mobile and portable RF communication devices and the
"KINETEC PERFORMA" unit
The "KINETEC PERFORMA" splint is designed to be used in an electromagnetic environment in which radiated RF interference is
controlled. The "KINETEC PERFORMA" splint customer or user can help prevent electromagnetic interference by maintaining a minimum
distance between mobile and portable RF communication devices (emitters) and the "KINETEC PERFORMA" splint, as recommended
below, according to the communication device's maximum output power.
Maximum assigned output
power for the emitter
W
0,01
0,1
1
10
100
For emitters whose assigned maximum emitted power is not given above, the recommended separation distance d in metres (m) can
be estimated using the equation applicable to the emitter frequency, where P is the emitter's maximum emission power characteristic in
watts (W) according to the latter's manufacturer.
NOTE 1
At 80 and 800 MHz, the separation distance for the highest frequency range is applicable.
NOTE 2
These directives cannot be applied in every situation. Electromagnetic propagation is affected by absorption and
reflection by structures, objects and people.
Performa
IEC 60601
Test level
3 Veff
from 150 kHz to 80 MHz
3 V/m
from 80 MHz to 2.5 GHz
from 150 kHz to 80 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
Compliance
level
Mobile and portable RF communication devices should not
be used closer to any part of the "KINETEC PERFORMA" splint,
including its cables, than the recommended separation
distance, calculated based on the equation applicable to
the emitter's frequency.
Recommended separation distance
3 V
P
d = 1.2
P
d = 1.2
3 V/m
P
d = 2.3
where P is the emitter's maximum output power characteristic
in watts (W), according to the emitter's manufacturer, and d is
the recommended separation distance in metres (m).
The field intensities of fixed RF emitters, determined by an on-
site electromagnetic investigation
compliance level in each frequency range
There may be interference near appliances bearing the
following symbol:
Separation distance according to the emitter's frequency
from 150 kHz to 80 MHz
P
d = 1.2
EN
Electromagnetic environment – guidance
from 80 MHz to 800 MHz
from 800 MHz to 2.5 GHz
, should be below the
a
m
from 150 kHz to 80 MHz
P
0.12
0.38
1.2
3.8
12
.
b
P
d = 1.2
0.12
0.38
1.2
3.8
12