Chlorine With Liquid Reagent - Lovibond Tintometer MD200 Mode D'emploi

Photometer-system
Masquer les pouces Voir aussi pour Tintometer MD200:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Methods
GB
CL 6
0.0.0
Zero
Test
CL 6
RESULT
Zero
!
Test
CL 6
RESULT
34

Chlorine with liquid reagent

0.02 – 4.0 mg/l
a) free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Remove the vial from the sample chamber and empty the vial.
Fill the vial with drops of the same size by holding the bottle vertically
and squeeze slowly:
6 drops of DPD 1 buffer solution
2 drops of DPD 1 reagent solution
Add the water sample to the 10 ml mark.
Close the vial tightly with the cap and invert several times to mix the
contents.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.
b) total Chlorine
Immediately after measurement add
3 drops of DPD 3 solution to the already coloured test solution.
Close the vial tightly with the cap and invert several times to mix the
contents.
Place the vial in the sample chamber making sure that the
are aligned.
Wait for a reaction period of 2 minutes.
(Countdown can be activated, see page 31)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.
c) combined Chlorine
combined Chlorine = total Chlorine – free Chlorine
Tolerances:
0 – 1 mg/l: ± 0.05 mg/l
> 1 – 2 mg/l: ± 0.10 mg/l
> 2 – 3 mg/l: ± 0.20 mg/l
> 3 – 4 mg/l: ± 0.30 mg/l
marks
marks
MD200_4 03/2014