5.6 Connect the tubes
Arrow symbols identify the inlet and outlet ports for the syringe and the pump connections. The "up"
arrow is the outlet port. The "down" arrow is the inlet port. Turn the tube connectors on the inlet and
outlet ports of the syringe and pump until they click.
The suction tube that drains into the waste bottle must be installed on the bottom of the beaker for
good drainage.
The diffusion tip and tube holder on the outlet tube, are pre-installed at the optimal positions. Do not
change the position of the diffusion tip or the tube holder.
5.7 Install the sensor
5.7.1 Connect the sensor
Use a conical adapter to hold the sensor tightly in the sensor holder. Make sure the conical adapter
is installed correctly. This makes sure of a secure water-tight fitting in the sensor holder and the tip of
the sensor will be in the correct position in the beaker.
Connect the sensor to an available sensor port on the rear of the instrument. After the sensor is
connected, make sure that the sensor icon shows in the banner at the top of the display.