The default speed for both aspiration and dispensing for each pipetting
range is as follows:
Pipetting
range
25 ml
10 ml
5 ml
Note: The speed can not be changed in the middle of the aspiration
or dispensing cycles.

5.3. Setting the Pipetting range (the pipette/tip size to be used)

To avoid accidental over aspiration the XL can be programmed to perform
on a selected range when using smaller than 25 ml pipettes/tips. There are
designated limits for 1, 2, 5, 10 and 25 ml tips. The selected limit remains
active for all the operating modes, e.g. in the C-mode the piston movement
stops if the aspiration goes beyond the preset volume limit, and in the P-,
d-, and Sd-modes the programming can be done only within the allowed
range.
1. Press
2. Press
3. Press
Note: The chosen volume range remains active for all the operating
modes unless readjusted.
Maximum
volume
1 ml
2 ml
5 ml
10 ml
25 ml
10
Default
speed
5
4
3
to display the current range.
or
until the desired range is displayed.
to confirm range selection.
Range
(P-mode)
0.1-1.0 ml
0.2-2.0 ml
0.5-5.0 ml
1.0-10.0 ml
2.0-25.0 ml
Pipetting
range
2 ml
1 ml
Range
(d-mode)
0.1-0.5 ml
0.1-1.0 ml
0.1-2.5 ml
0.5-5.0 ml
1.0-12.5 ml
Default
speed
2
1
Range
(Sd-mode)
0.1-1.0 ml
0.1-2.0 ml
0.1-5.0 ml
0.5-10.0 ml
1.0-25.0 ml
Range
(C-mode)
0-1.0 ml
0-2.0 ml
0-5.0 ml
0-10.0 ml
0-25.0 ml