I-Tech LymphoPress 4 Mode D'emploi page 22

Pressothérapie modèle
Table des Matières

Publicité

Electrical
fast
transient/burst
IEC 61000-4-4
Impulses
EN 61000-4-5
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
Mains
power
electromagnetic field
EN 61000-4-8
IACER SRL
± 2kV for power
supply lines
±1kV
differential
mode
< 5% U
T
(>95% dips of U
)
T
per 0,5 cycles
40% U
T
(60% dips of U
)
T
per 5 cycles
70% U
T
(30% dips of U
)
T
per 25 cycles
< 5% U
T
(>95% dips of U
)
T
per 5 seconds
3 A/m
Page 22 de 28
± 2kV per power
supply lines
±1kV
differential
mode
< 5% U
T
(>95% dips of U
)
T
per 0,5 cycles
40% U
T
(60% dips of U
)
T
per 5 cycles
70% U
T
(30% dips of U
)
T
per 25 cycles
< 5% U
T
(>95% dips of U
)
T
per 5 seconds
3 A/m
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
If the user of the
LYMPHOPRESS 4
requires
continued
operation
during power mains
interruptions,
it
is
recommended
that
LYMPHOPRESS 4
be powered from an
uninterruptible
power supply or a
battery.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
MNPG93-01 del 05/05/15