Eaton MGR10 Serie Manuel D'utilisation page 76

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

*RST
Clears all pending operations, and resets the instrument to the default RESET condition as
described in the instrument user manual with two exceptions:
1) No change is made to the remote interface parameters (interface type,
rate and IEEE address).
2) The trigger mode is set to SINGLE (INITiate:CONTinuous OFF).
For IEEE only, this also resets the Operation Complete Command State and Operation
Complete Query State (clears bit 0 of Standard Event Register).
The default conditions that are forced are:
Range:
Current:
Read Rate:
Trigger Mode:
Measurement Limits:
Filter:
Settling Algorithm:
Open Circuit Limit:
Temp Compensation:
Data Logger:
READ? or FETCh?:
Talk Only mode: OFF
Analogue Output OFF
*TST?
Always returns '0' to indicate self-test OK.
*WAI
This command is accepted by both interfaces, but ignored as all commands are executed
sequentially. It is provided only for compatibility with IEEE-488.2.
30KOHM, AUTO1
100%, +I
SLOW
CONT:OFF (Single), STOP
OFF, Alarm ON, Lower Limit = 0, Upper Limit = 30K
OFF, 10 readings
OFF, 10 readings, 10 digits limit
OFF
OFF, MAN, +20C, Cu, USER = 3980, Ref Temp = +20C
OFF, Count = 10, Datalog Memory unchanged.
Set to FRESistance.
MGR10 A/B/C Indice : J (version 7.0 du logiciel)
baud
76

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Mgr10Mgr10aMgr10bMgr10cSefelec mgr10 serieSefelec mgr10 ... Afficher tout

Table des Matières