Interface Functions - Eaton MGR10 Serie Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6.2.2.3 Interface Functions

The interface performs the following functions;
SH1, AH1, T5, TE0, L3, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2
These have the following meanings :
SH1
Source Handshake
The device can send multiline messages.
AH1
Acceptor Handshake
The device can receive multiline messages.
T5
Talker Capability
The device;
;
;
;
TE0
Extended Talker Capability
Extended talker addressing is not implemented.
L3
Listener Capability
The device;
;
;
LE0
Extended Listener Capability
Extended listener addressing is not implemented
SR1
Service Request Function
The device can asynchronously assert the SRQ line to request the attention of the system
controller.
RL1
Remote/Local Function
The device has full remote/local switching capability. There is no manual switch on the
instrument.
PP0
Parallel Poll Function
The parallel poll function is not implemented.
DC1
Device Clear Function
The device may be set to the default state by the appropriate command, and includes a
selective device clear.
can act as a basic talker
can respond to a serial poll
can perform a talk only mode
unaddresses if MLA sent
can act as a basic listener
can be set to listen only mode
becomes unaddressed if MTA sent
MGR10 A/B/C Indice : J (version 7.0 du logiciel)