Adjust The Display Contrast; Adjust The Temperature - HACH LANGE sensION+ EC71 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 48
3. Use the
or
to select
Option
Description
Deactivated
Select Deactivated if no printer or PC is connected.
For Printer
Select Dot matrix printer or Thermal printer.
For Computer Select Terminal, LabCom or LabCom Easy. The LabCom
Software controls several modules, pH and conductivity
meters, automatic burettes, Samplers and so on from a
computer. The LabCom Easy software gets pH and
conductivity data from a PC.
Change the date and time
The date and time can be changed from the Date / Time menu.
1. From the main menu use the
2. Use the
or
to select Date / Time and confirm.
3. Use the and to advance to the next step and use the
change a value. Confirm.
The current date and time will be shown on the display.

Adjust the display contrast

1. From the main menu use the
2. Use the
or
to select Display contrast and confirm.
3. Use the and to adjust the contrast of the display and confirm.

Adjust the temperature

The temperature measurement can be adjusted at 25 °C (77 °F) and/or
85 °C (185 °F) to increase accuracy.
1. Put the probe and a reference thermometer in a container of water at
approximately 25 °C and allow the temperature to stabilize.
2. Compare the temperature read by the meter with that of the
reference thermometer. The difference is the adjustment value for
the meter.
10 English
or
to select SYSTEM. Confirm.
or
or
to select SYSTEM. Confirm.
Example: reference thermometer: 24.5 °C; meter: 24.3 °C.
Adjustment value: 0.2 °C.
3. Enter the adjustment value for the 25 °C reading:
a. From the main menu use the
b. Use the
or
c. Use the
or
d. Use the arrow keys to enter the adjustment value for 25 °C.
Confirm.
4. Put the probe and a reference thermometer in a container of water at
approximately 85 °C and allow the temperature to stabilize.
5. Compare the temperature from the meter with that of the reference
thermometer. The difference is the adjustment value for the meter.
a. Use the
or
b. Use the arrow keys to enter the adjustment value for 85 °C.
Confirm.
c. Select Save changes and confirm.
to
Change the stirring settings
The magnetic stirrer can be turned on and the stirring speed can be
changed in the Stirring menu.
1. From the main menu use the
2. Use the
or
to select Stirring and confirm.
3. To turn on/off the stirrer push
4. When the stirrer is turned on, use the
speed in %.
Note: Use the
or
during a measurement.
Change the temperature units
The temperature units can be changed to Celsius or Fahrenheit.
1. From the main menu use the
2. Use the
or
to select Temperature units and confirm.
or
to select SYSTEM. Confirm.
to select Readjust temp. and confirm.
to select 25 °C and confirm.
to select 85 °C and confirm.
or
to select SYSTEM. Confirm.
.
or
to change the stirring
to change the stirring speed during calibration and
or
to select SYSTEM. Confirm.