Customer Service; Technical Data - uebe Visomat double comfort Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

F

2. Customer service

The unit may only be repaired by the manufacturer or by an expressly authorised
organisation.
Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Phone:
+49 (0) 9342/924040
Fax:
+49 (0) 9342/924080
E-mail:
info@uebe.com
Internet:
www.uebe.com

3. Technical data

Unit type:
Digital automatic unit with electric pump for measur-
ing blood pressure on upper arm
Dimensions:
L = 115 mm x W = 150 mm x H = 54 mm
Weight:
330 g not including batteries
Display:
LCD display (liquid crystal display) for measured val-
ues and check displays
Memory:
2 memories each with 60 measured values (stored
automatically) and average value (A)
Measurement pro-
Combines auscultatoric and oscillometric determina-
cedure:
tion of systole, diastole and pulse
Reference procedure
Auscultatoric measurement
of clinical testing:
Inflating pressure:
approx. 30-40 mmHg above systolic blood pressure
Pressure display range:
0-300 mmHg
EN-82
Technical information
Technical information
Measurement range
Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse: 40-160 pulses/min
Error limits:
Blood pressure measurement: corresponds to EN
1060 Part 3
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5%
Serial number:
The unit bears a serial number SN which provides
clear identification.
Power supply:
Battery type: 4 x 1.5 V AA alkali manganese (LR 6) or
lithium (FR 6) round cells, service life: more than 800
measurements in 2 years
Optional: Stabilised type A1 mains adapter, output 6
VDC, minimum 600 mA
Cuff with microphone:
Metal ring cuff for arm circumferences of 23-43 cm,
type UWK
Operating conditions:
Ambient temperature 10 to 40 °C, Rel. air humidity
up to 85% not condensing,
Air pressure 700 to 1040 hPa
Storage and transport
Ambient temperature -5 to +50 °C,
conditions:
Relative humidity up to 85%, non-condensing
Air deflation valve:
Electronically controlled
Automatic switch-off
approx. 3 minutes after end of measurement
Radio clock signal:
DCF77
Protection class IP:
IP 20: Protected against solid foreign particles with
a diameter of more than 12.5 mm, no protection
against water.
Expected service life:
5 years
Classification:
Internal power supply by battery
F
EN-83