Factory calibration
Measured value
Important: For water activity, setting
the offset changes the angle of the
characteristic with zero as centre point.
page 60
Additional information for one point calibration of temperature
(Channel/Sensor 1):
CAL1 initiates the one point calibration for temperature. The actual value
of the temperature (Channel/Sensor 1) is shown in the upper half of the
display. The lower half displays the temperature offset, which can be
adjusted per ±10°C (±18°F)in steps of 0.1°C. The factory settings can be
restored by setting the temperature offset to 0.0°C.
Additional information for one point calibration of water activity (Channel/
Sensor 2):
CAL2 initiates the one point calibration for the water activity. The actual
value (Channel/Sensor 2) is shown in the lower half of the display. The
Offset
sensor offset value, which can be adjusted per +0.100 in
0.001°C steps, flashes in the upper half of the display. The factory
New calibration
line
settings can be restored by setting the offset to 0.000.
The calibration point should be in the range 0.3...0.95a w (recommended
0.76 a w ).
Important: For water activity, setting the offset changes the angle of the
characteristic with zero as centre point.
OMNIPORT 20 english