Manual
Important: Calibration should only be
carried out by skilled personnel using
suitable calibration standards.
page 44
One point calibration of SDI sensors such as combination
probe for humidity / temperature.
CAL1: CAL1 initiates the one point calibration to channel 1 (Sensor 1).
The symbols for Sensor 1 and Sensor 2 disappear from the
display. The actual value for channel 1 (Sensor 1) is now displayed
in the upper half of the display. The offset value for channel 1
(Sensor 1) flashes in the lower area; this value can be incremented
using
and decremented using
with
.
The menu is than exited automatically and CAL END is displayed.
Incorrect calibration is indicated by CAL FAIL and must be repeated.
The factory settings can be restored by setting the offset to 0.0.
OMNIPORT 20 english
. The offset value is confirmed