Installation - Masterflex L/S 07519-05 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Tool Required: Allen Key (provided).
Mount the Pump Head to a MASTERFLEX
1. Install the plastic Boot over the tang at the end of pump shaft, (Figure 1). This Boot prevents metal-to-metal contact,
and thus reduces noise. This Boot should be inspected and replaced when worn or at 500 hour intervals.
CAUTION:
Not using the boot will damage the pump.
NOTE: The boot with closed ends (A-3480) will fit most drives. Use of the boot with open ends (A-3514) may be required
on some drives. Both boots are supplied with the pump.
2. Connect the Pump to Drive by aligning the tang on Pump Head, (including installed boot) with the slot in the motor
Drive shaft.
3. Attach Pump Head to drive unit with four screws (provided) using the Allen Key (also provided).
WARNING: Tubing breakage may result in fluid being sprayed from pump. Use appropriate measures to pro-
tect operator and equipment.
4. If the Pump Head requires support, install the appropriate length support screw with plastic foot and locknut (three
provided). Level the Pump Head with the drive unit by adjusting the foot. Tighten the locknut against the bottom of the
pump body.

INSTALLATION

®
®
L/S
Pump Drive as follows:
FIGURE 1 PUMP MOUNTING
A-3480
FIGURE 2 PLASTIC BOOTS
4
A-3514

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

L/s 07519-06

Table des Matières