Setup; Select Tubing; Tube Loading - Masterflex L/S 07534-04 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Support foot
Figure 3. Stacking Pump Heads
Adjustable support foot may be used to stabilize pump stack.
Rotate the foot on the front pump as required to prevent the
pump and drive from rocking.

4. SETUP

Multichannel Pump Heads accommodate a broad range of
tubing sizes as noted in Section 5.1.

4.1 SELECT TUBING

Depending on which Pump Head is used, up to eight tubing
sets can be teamed up in a single Pump Head. Tubing sizes
can be intermixed within each occlusion bed as long as the
same tubing material is used.
Use only MASTERFLEX two-stop tubing sets (Figure 4)
withMASTERFLEX multichannel tubing pumps to ensure
optimum performance. Use of other tubing may void
applicable warranties.
Tubing
®
Figure 4. Masterflex
Two-Stop Tubing Set
4
Collar
Collar

4.2 TUBE LOADING

WARNING: Turn off drive before removing
or installing tubing. Safety guard is provided
to minimize risk of fingers getting caught
between roller mechanism and the base of
the module. However, be safe—keep your
fingers away from these areas.
CAUTION: Use only tubing sets listed in
Appendix B. Use of other tubing could cause
malfunction or damage to the pump.
With the drive stopped, follow this procedure to install tubing:
1. Release latch and open occlusion bed (Figure 5).
2. From the top of the Pump Head, insert one end of the
tubing over rollers. While holding occlusion bed, open spring
loaded safety guard enough so that tubing can be inserted
between guard and slotted tubing retainer (Figure 6).
3. Place tubing stop collar into slot to secure (Figure 7).
4. Making sure not to twist tubing, lightly stretch it across
rollers to secure tubing stop collar in corresponding slot on the
opposite end of Pump Head (latch side) (Figure 8).
Repeat the process for additional tubing channels as required.
Figure 5. Release latch and open occlusion bed.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

L/s 07534-08L/s 07535-04L/s 07535-08L/s 07536-02L/s 07536-04

Table des Matières